Product Code: LBNY-0001-500012
Categories: Analytical Instrument
High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD)
High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD)
General Features:
Analyses of mycotoxins, PAHs, pesticides formulation etc .
High Resolution Liquid Chromatograph:
Quaternary pump – 1 pc.
Flow: 0.001–5 mL/min, in 1 μL/min increments .
Pressure: 20 - 700 bar with the flow rate up to 5 mL/min .
Serial dual-piston pump
mixing up to 4 solvents at low pressure .
Compressibility Compensation: Fully automated, independent of mobile phase composition
Leak detection included
Vacuum degasser: Built-in, with 4 channels .
Flow precision: <0.07% RSD .
Flow accuracy: ± 1% .
Proportioning precision: <0.15% RSD .
Proportioning Accuracy: ±0.5% of full-scale .
Autosampler – 1 pc.
Injection volume range: from 0.1 to 100 μl .
Precision of injection: <0,25% RSD .
Injection volume accuracy: ±0.5% .
Pressure range: 20 - 700 bar .
Injection Linearity: r >0.99999
Sample capacity: more than 210 vials of 1,5/2 ml volume
Carry over: 0.003% .
Temperature range: from 4 °C to 40 °C .
Temperature stability: ± 1 °C .
Leak detection
Thermostated column compartment – 1 pc.
Work principle: Still air
Temperature range: from +10 °C to 85 °C .
Temperature stability: ± 0.05 °C .
Temperature accuracy: ±0.5 °C .
Column capacity: 2 columns, up to 300 mm length .
leak sensors
DAD detector – 1 pc.
Optical design: 1024 element photodiode array
Light source: Deuterium lamp
Wavelength range: 190-800 nm
Wavelength accuracy: ±1 nm .
Wavelength repeatability: ±0.1 nm .
Number of signals: 8 + 3D .
Scanning rate: 125 Hz
Noise: <±6 μAU .
Drift: <1 mAU/h .
Stainless steel flow cell: 3 μL or less, 12 MPa ., 10 mm optical path length
Leak sensor
FLD – 1 pc.
Optical design: Two monochromators with concave holographic gratings and elliptic mirrors
Light source: Xenon flash lamp
Excitation wavelength range: 200–630 nm .
Emission wavelength range: 200–650 nm .
Spectrum scanning modes: Single Spectrum Scans or FL Field Acquisitions: Excitation, emission or synchronous mode .
Wavelength accuracy: ±2 nm .
Wavelength repeatability: ±0.2 nm .
Wavelength calibration: Internal calibration, excitation monochromator with emission lines of xenon flash lamp, emission monochromator with Raman shift of water and emission lines of xenon lamp .
Number of signal channels: up to 4 .
Data collection rate (single-channel): Up to 200 Hz .
PC and monitor.
Software for data collection, processing, and reporting of results.
High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD), High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) Equipments, High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) Tools, High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) Tool Kits, High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) Manufacturers, High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) Suppliers from India, China, Kenya
Similar items like High Performance Liquid Chromatograph With Diode Array Detector And Spectrofluorimeter Detector (HPLC DAD SFD) you may view