Oxygen concentrators are not suitable to be used as an oxygen source to other medical equipment providing respiratory support, such as ventilators, CPAP devices, High Flow Nasal Oxygen, etc.
Provides a continuous, variable flow of concentrated oxygen 93% (±3%) derived from room air, to a maximum rate 8 L/min.
Equipped with a single oxygen outlet provided with controllable flowmeter.
Audible and/or visual alarms for low oxygen concentration (<82%) and power supply failure.
Contains oxygen monitor to verify concentration.
Flowmeter minimum flow rate of 2 L/min. Lower flow rates may be achieved by a combination of the concentrator and a separately supplied flow splitter (details under RELATED PRODUCTS).
Audible and visual alarms for high temperature and pressure failure.
Power efficiency ≤ 70 W/LPM.
Digital meter that displays cumulative hours of device operation.
Oxygen outlet(s) with 6 mm (¼-inch) barbed fitting.
Flowmeter continuously adjustable, with markings of at most 0.5 LPM intervals.
Noise level <60 dB(A).
CASING AND ENVIRONMENT
Oxygen outlet to be securely mounted and sheltered.
The unit includes internally and externally mounted filters for cleaning the air intake.
All user-removable filters are cleanable. Cleaning instructions for filters are included in the instructions for use.
Case to be hard, to be cleanable with standard hospital cleaning materials.
Whole unit to be movable with four wheels.
ELECTRICAL CHARACTERISTICS
Electrical power requirements to be locally compatible.
Capacity for safe operation from at least ± 10% of rated voltage.
Electrical protection by resettable circuit breakers or replaceable fuses.
230 Volt, 50 Hz, 2.1 amps (110 Volt units available upon request)
Mains power cable to have length ≥ 2.5 m.
OXYGEN PRODUCTION
Provides a continuous, variable flow of concentrated oxygen 93% (±3%) derived from room air, to a maximum rate 8 L/min.
Audible and visual alarms for high temperature and pressure failure.
Power efficiency ≤ 70 W/LPM.
Digital meter that displays cumulative hours of device operation.
Oxygen outlet(s) with 6 mm (¼-inch) barbed fitting.
Contains oxygen monitor to verify concentration.
Equipped with two oxygen outlets each provided with controllable flow meter.
Audible and/or visual alarms for low oxygen concentration (<82%) and power supply failure.
Flowmeter continuously adjustable, with markings of at most 0.5 LPM intervals.
Noise level <60 dB(A).
Flowmeter minimum flow rate of 2 L/min. Lower flow rates may be achieved by a combination of the concentrator and a separately supplied flow splitter.
ELECTRICAL CHARACTERISTICS
Mains power cable to have length ≥ 2.5 m.
Capacity for safe operation from at least ± 10% of rated voltage.
Electrical power requirements to be locally compatible.
230 Volt, 50 Hz, 2.1 amps (110 Volt units available)
Electrical protection by resettable circuit breakers or replaceable fuses.
CASING AND ENVIRONMENT
Case to be hard, to be cleanable with standard hospital cleaning materials.
All user-removable filters are cleanable. Cleaning instructions for filters are included in the instructions for use.
Oxygen outlet to be securely mounted and sheltered.
Whole unit to be movable with four wheels.
The unit includes internally and externally mounted filters for cleaning the air intake.