Search

Search by Alphabet :

A B C D E F G H I J K L M N O P Q R S T U V W X Y Z

Sensor, reusable, Rad-G, neonate, YI set

Product Code: LBNY-0017-2200039 - (Sensor, reusable, Rad-G, neonate, YI set)

The Reusable Sensor is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

....


   

Resuscitator, upright, neonate, set

Product Code: LBNY-0017-2200042 - (Resuscitator, upright, neonate, set)

Hand-operated in a vertical position; the inflatable bag is operated an upright position perpendicular to the body being ventilated.
The resuscitator is autoclavable between 121°C and 134°C.
The resuscitator has a provision which allows connection of oxygen tubing.
Portable, self-inflating, reusable.
The resuscitator is equipped with a non-rebreather valve.
Resuscitator is easy to disassemble, clean, disinfect, and reassemble.
The capacity of the self-inflating ventilation bag is between 220 ml.
The resuscitator is equipped with a pressure release (limiting) valve which safeguards against generating ventilation pressures over 45 cmH?O, while still allowing for an airway pressure of at least 30 cmH?O.
All parts manufactured from high-strength, long-life materials that require no special maintenance or storage conditions.
The resuscitator, all its components and accessories are free of natural rubber latex.
The pressure release valve should allow to be overridden by the user when higher pressures need to be generated in the airways.
When the pressure release valve is not overridden by the user, it should automatically return to its default position whereby it protects against excess pressures. Users should not be able to lock the pressure release valve in a permanent closed position whereby there is no longer a safeguard against excessive ventilation pressures.
Resuscitators are used for ventilation assistance in critical situations of respiratory distress or failure. It is considered essential equipment for health structures and emergency rescue team (to be available in emergency room, delivery room, operating theatre, intensive care unit, ambulance, etc.).

....


   

Sensor, reusable, pulse ox., neonate

Product Code: LBNY-0017-2200049 - (Sensor, reusable, pulse ox., neonate)

The reusable sensors are indicated for spot-checking or continuous, non-invasive measurement of arterial oxygen saturation (SpO2) and pulse rate for use with neonates in operating rooms, recovery rooms, intensive care units (ICUs) and wards.
Design: Reusable, Y-style with built-in wrap-around belt for use on big toe or thumb.
Nellcor standard compatible sensor.
Cable length: 3.0m
Plug type: DB9 pin
Patient weight: 3 - 20kg.

....


   
Support