Finger Tip Pulse Oximeter
....Handheld, portable, battery powered pulse oximeter, continuously displaying patient oxygen saturation (SpO2) and pulse rate in real time using an external probe on the skin.
Display:
Display with main parameters: SpO2, pulse rate, plethysmographic waveform, signal quality, alarm messages, battery state indication.
Data update period for valid data displayed ≤ 15s.
Features:
Audible and visual alarms for low/high saturation and pulse rate.
Audible and visual alarms for sensor error or disconnected, system errors, low battery.
Suitable for cleaning and disinfection with hospital-grade cleaning products.
Alarm override and temporary silencing function.
Capable of working with adult, paediatric and neonatal reusable probes.
Enclosure protection IPX1 or better.
Device and probe weight < 200 g.
Measurement:
SpO2 and pulse rate monitor, with plethysmography waveform, for adults, children and neonates.
SpO2 resolution 1 %.
SpO2 accuracy (in the range 70 - 100%) within ± 3%, for all patients and perfusion / movement conditions.
SpO2 detection includes the range 70 - 100 %. Also capable of measuring lower saturations.
Pulse Rate detection includes the range 30 - 240 bpm.
Pulse Rate accuracy within ± 2 bpm or ± 2%.
Pulse Rate resolution 1 bpm.
Handheld, portable, rechargeable battery powered pulse oximeter, for the noninvasive spot-checking or continuous monitoring of patient oxygen saturation (SpO2), pulse rate and respiratory rate.
Pulse Rate detection range: 25 - 240 bpm.
Pulse Rate resolution: 1 bpm.
Pulse Rate accuracy: within ± 3 bpm (no motion, low perfusion), ± 5 bpm (motion) for all patients (infants, children, adults).
Respiratory rate range: 4 to 70 rpm
Respiratory rate resolution: 1 rpm
Respiratory rate accuracy (no motion, adults/paediatrics (> 2 years)): within ± 3 rpm A(RMS), ± 1 rpm mean error
(Note: device is indicated for spot-checking or continuous monitoring of Respiration Rate from the photoplethysmogram (RRp) of adult and pediatric patients (> 2 years) during no motion conditions).
Measurement:
SpO2 and pulse rate monitor, with plethysmography waveform, for infants, children and adults.
SpO2 detection includes the range: 70 - 100 %.
SpO2 resolution: 1 %.
SpO2 accuracy (in the range at least 70 - 100%): ± 2% (no motion, low perfusion), ± 3% (motion) for all patients (infants, children, adults).
Display:
Data display refresh rate: every 1 s
Display with main parameters: %SpO2, pulse rate, respiratory rate, signal quality, status messages, battery state indication.
Casing:
Enclosure protection IP22.
Case and reusable probe suitable for cleaning and disinfection with hospital-grade cleaning products.
Overall device and probe weight < 300 g.
Electrical characteristics:
Operated by rechargeable, non-replaceable Li-ion battery power supply.
External AC battery charger, plug style and voltage as per local supply.
Charger certified to IEC 60601-1.
Protection against defibrillator discharges.
Suitable for operating by mains power and by battery.
Automatic switch between battery and mains powered modes.
The display shows which power source is in use.
Running time on battery only ≥ 12 hours (24h at lowest screen brightness and beep tone turned off).
Features:
Suitable for detection in low perfusion conditions.
Design enables use in demanding environments, e.g. shock, vibration.
Automatic correction for movement and ambient light artefacts.
Audible and visual alarms for low saturation and high/low pulse rate, thresholds settable by user.
Audible and visual alarms for sensor error or disconnected, system errors, low battery.
Alarm override and temporary silencing function.
Capable of working with, and supplied with, adult, paediatric and neonatal reusable probes.
Handheld, portable, rechargeable battery powered pulse oximeter, displaying patient oxygen saturation (SpO2), pulse rate and respiratory rate in real time.
Features:
Suitable for detection in low perfusion conditions down to perfusion index of 0.1%.
Design enables use in demanding environments, e.g. shock, vibration
Automatic correction for movement and ambient light artefacts
Audible and visual alarms for sensor error or disconnected, system errors, low battery.
Measurement:
SpO2, pulse rate and respiratory rate monitor, for infants, children, adults.
SpO2 detection includes the range: 70 - 100 %.
SpO2 resolution: 1 %.
SpO2 accuracy (in the range at least 70 - 100%): ± 2% (no motion, low perfusion), ± 3% (motion) for all patients (infants, children and adults).
Pulse Rate detection includes the range: 25 - 240 bpm.
Pulse Rate resolution: 1 bpm.
Pulse Rate accuracy: within ± 3 bpm (no motion, low perfusion), ± 5 bpm (motion) for all patients (infants, children, adults).
Respiratory rate includes the range: 4 to 90 rpm.
Respiratory rate resolution: 1 rpm
Respiratory rate accuracy within ± 1 rpm in ideal conditions
Electrical characteristics:
Operated by rechargeable, non-replaceable Li-ion battery, power supply
External AC battery charger, plug style and voltage as per local supply.
Charger certified to IEC 60601-1.
Protection against defibrillator discharges.
Suitable for operation by battery and by mains power supply, if connected and/or recharging.
Automatic switch between battery and mains powered modes, when recharging or in mains power failure.
The display shows which power source is in use.
Running time on battery only ≥ 12 hours (24h at lowest screen brightness and beep tone turned off).
Automatic power-off after period of no use.
Display:
Data display refresh rate ≤ 1 s
Display with main parameters: %SpO2, pulse rate, respiratory rate, signal quality, status messages, battery state indication.
Patient age group input feature.
Display indicates high and normal respiratory rate count for inputted age.
Display indicates when SpO2 < 90% (hypoxaemia per WHO guidelines.)
Casing:
Enclosure protection IPX2.
Case and probe suitable for cleaning and disinfection with hospital-grade cleaning products.
Overall device and probe weight < 300 g.