Suitable for use in resource-limited settings. For the separation of heterogeneous mixtures or samples in clinical laboratory applications.
....Suitable for professional clinical use, all patient categories neonate, infant, adult
Robust shock resistant design allow use in demanding environments
Ultra-compact pulse oximeter integrated into finger/toe clip
Measuring ranges:
SpO2 resolution: 1% or less
SpO2 accuracy within: +/-3%
SpO2 detection within the range of: 70 to 99%
Pulse rate accuracy: within: +/-3%
Pulse rate detection: 30 to 240 bpm
Pulse rate resolution: 1 bpm)
Enclosure protection IPX2 or better
Suitable for cleaning and disinfection with hospital-grade cleaning products.
Display shows SpO2, pulse rate, signal strength, sensor error or disconnect and low battery.
Power requirements two 1.5V AAA batteries
Accommodates finger/toe thicknesses 8 to 25mm directly or through the use of a separate sensor which can be connected to the device.
Spot-check of arterial blood oxygen saturation (SpO2) and heart rate(HR)
Batteries must allow at least 2500 spot-checks, calculated at 30 seconds per spot-check, or at least 21 hours of operation.
The unit is equipped with an automatic power off feature.